MICARE
CODE: NITR-GLOVE-S/M/L/XL
There is an undeniable difference when you put on MICARE Nitrile Examination Gloves. The outstanding comfort and protection gives you the confidence you need to perform your daily tasks.
Features
Aesthetically Presentable
Unlike other gloves, our unique glove packing method eliminates glove creases – giving you an ironed finish.
Finger Tip Textured
All our nitrile gloves are finger textured for that added handling precision.
Proven Protection
Gloves are clinically tested to handle against a wide range of chemicals which also comply with food safety standards.
Hygienic
Our fully automated glove production process eliminates any possible human contamination from production down to packing.
Comfortable & Stretchy
The ability of our gloves to not only expand, but to also retract according to your hand movements rewards you with a free range motion.
Extended Version
Longer cuff available for that extra protection against splashes over the wrists and arms.
Beaded Cuff
Increased strength & ease in donning.
Technical Info
How GoodpacTM technology dispenses one glove at a time
With an alternating layering method, each top layer glove cuff will always completely cover the fingers of its immediate bottom layer glove. Hence, exposing only the top most layer for the wearer to pull. Therefore preventing accidental pulling of the bottom layered gloves. This results in zero wastage. Moreover, a neatly packed glove box will prevent creased gloves therefore speeding up the process of dispensing and donning.
Technical Info
MICARE Glove Technical Specification
Product conforms to the following test standards:
ASTM D6319 | Standard Specification for Nitrile Examination Gloves | |
ASTM D412 | Tensile Properties of Vulcanised Rubbers | |
ASTM D573 | Accelerated Aging for Latex (at 70°C for 166+-2hr) | |
ASTM D5151 | Standard Test for Detection of Holes in Medical Gloves | |
ASTM D6124 | Standard Test Method for Residual Powder on Medical Gloves | |
ISO 2859-1 | Sampling Procedures for Inspection by Attributes Part 1: Sampling Schemes Indexed by Acceptance Quality Limit (AQL) for Lot-by-lot Inspection |
Performance Requirements
CHARACTERISTIC | SAMPLING PLAN | INSPECTION LEVEL | AQL |
---|---|---|---|
Watertight Test | Multiple Normal | G1 | 1.5 |
Visual Test | Single Normal | S4 | Major 2.5 / Minor 4.0 |
Physical Dimension | Single Normal | S2 | AQL 4.0 |
Physical Properties | Single Reduced | S2 | AQL 4.0 |
Particulate Residue | N/A | N=5 | <2mg glove |
Physical Dimensions
SIZE | EX SMALL | SMALL | MEDIUM | LARGE | EX LARGE |
---|---|---|---|---|---|
Palm width ± 4 mm | 76 | 86 | 98 | 107 | 115 |
Glove length (mm) | Min 230 |
Physical Properties
TEST | BEFORE AGING | AFTER AGING |
---|---|---|
Tensile Strength (MPa) | Min 18.0 | Min 16.0 |
Elongation (%) | Min 500 | Min 400 |